Yes. The DragLab Drying Oven Series includes a temperature calibration function that allows users to compensate for minor temperature deviations by entering a correction value through the controller when calibration is required.
This feature helps improve the agreement between the displayed chamber temperature and a calibrated external reference thermometer, supporting greater measurement accuracy for applications where precise temperature control is important. Temperature calibration is particularly useful after periodic maintenance, during routine quality assurance procedures, or when compliance with laboratory calibration protocols is required.
The calibration adjustment is performed through the controller interface and enables users to fine-tune the displayed temperature without affecting the oven's overall operating functions. For the most reliable results, DragLab recommends performing calibration using a certified reference thermometer under stable operating conditions and in accordance with the laboratory's internal calibration procedures.
Before cleaning the DragLab Drying Oven Series, switch off the oven, disconnect it from the power supply, and allow it to cool completely. The AISI 304 stainless steel chamber, shelves, and interior surfaces should be cleaned using a soft, lint-free cloth and a suitable non-corrosive, non-abrasive cleaning agent to remove dust, residues, or other contaminants.
Avoid using steel wool, abrasive pads, strong acids, alkalis, or aggressive solvents that could damage the stainless-steel surfaces or protective finishes. During cleaning, take care to prevent water or cleaning liquids from entering the controller, electrical components, or ventilation openings. After cleaning, ensure that all surfaces are completely dry before reconnecting the oven to the power supply and resuming operation.
Regular cleaning helps maintain a hygienic working environment, minimizes the risk of sample contamination, and contributes to the reliable long-term performance of the oven.
No. Water or other cleaning liquids should never be sprayed directly onto the DragLab Drying Oven Series. Moisture entering the controller, electrical components, or ventilation openings may create electrical hazards, cause equipment damage, and affect the oven's safe and reliable operation.
Instead, disconnect the oven from the power supply, allow it to cool completely, and clean the interior and exterior surfaces using a soft cloth lightly dampened with a suitable non-corrosive cleaning agent. After cleaning, ensure that all surfaces are completely dry before reconnecting the oven to the power supply and returning it to service.
DragLab recommends performing preventive maintenance at least once per year to ensure safe operation, maintain optimum heating performance, and extend the service life of the oven. Regular maintenance helps verify that key components, such as the heating system, air circulation system, temperature sensor, door seal, and electrical connections, are functioning correctly.
In laboratories where the oven is used continuously or under demanding operating conditions, more frequent inspections may be appropriate. Routine preventive maintenance can help identify potential issues before they affect performance, reduce unexpected downtime, and ensure reliable and consistent operation throughout the oven's service life.
No. The door gasket should only be replaced after the DragLab Drying Oven Series has been switched off, disconnected from the power supply, and allowed to cool completely. Replacing the gasket while the oven is still hot may damage the sealing material, reduce its sealing performance, or increase the risk of personal injury.
Before installing a new gasket, ensure that the gasket groove and door sealing surface are clean and free of debris. Correct installation of the gasket helps maintain an effective door seal, minimizes heat loss, and supports stable chamber temperatures and reliable heating performance.