| Working temperature range | min. +20°C ambient up to +100°C |
| Temperature fluctuation | 0.1°C |
| Temperature accuracy | ± 0.3 % (Over full scale) |
| Temperature unit | Easy to choose Celsius or Fahrenheit |
| Temperature sensor | 1 Pt100 sensor DIN class A in 2-wire-circuit |
| Language setting | English |
| Control | Adaptive multifunctional digital PID-microprocessor controller |
| Display | Initiative Touch Screen display with user-friendly software |
| Timer | Digital backward counter to target time setting |
| Timer control | Adjustable from 1 minute to 99:59 hour or continues |
| Delay timer | Programmable time delay function |
| Function set point wait | The timer does not start until the set temperature reached the set point |
| Adjustable parameter | Temperature and time, automatic fan adjustable |
| External door | External steel door with compression door lock |
| Internal door | Inner door made of tempered safety glass |
| Calibration certificate | Calibration certificate at +100°C (optional) |
| Interior | Easy-to-clean interior made of stainless steel AISI 304 on all sides |
| Gasket | Synthetic rubber gasket |
| Housing | Electrostatic powder-coated galvanized steel |
| Insulation | Fully isolated chamber by Rockwool covered by an aluminum sheet to ensure high-temperature stability |
| Emergency knob | Manual knob in case of emergency |
| External dimensions | 470 (W) X 525 (D) X 600 (H) mm |
| Shelves | 2-shelves chrome plated |
| Auto diagnostic system | Audible alarm for fault analysis |
| Alarm | Audible, can be switched off, adjustable limit on the independent temperature safety device |
| Standards | Integrated independent temperature safety device with visual alarm accordance DIN 12880 |
| Standards | EN 61010-1 and EN 61010-2-010 in accordance with EU directive 2014/35/EU |
| Standards | EN 61326-1:213 and EN 61326-1 in accordance with EU directive 2014/30/EU |
| Class | Class I in accordance with the EU directive MDD 93/42/EEC |
| Protection Class | IP20 according to EN 60529 |
| Internal dimensions | 320 (W) X 310 (D) X 320 (H) mm |
| Chamber | Easy to clean Stainless steel material no.: EN 1.4301 (AISI 304) |
| Useful volume | 30 liter |
| Standard shelves | 2-shelves chrome plated |
| Max. number of shelves | 5 |
| Permitted load | 75 Kg |
| Load per shelf | 15 Kg |
| Fan | Forced air circulation by quiet air turbine automatically adjustable |
| Installation | The distance between the wall and the rear of the appliance must be at least 15 cm. |
| Installation | The sides distance must be not less than 5 cm. |
| Installation | The distance from the ceiling must not be less than 20 cm |
| Temperature | +5 °C to +55 °C |
| Humidity | Max. 80% rh |
| Voltage | 230 volts, 50/60Hz |
| Watt | 1500 watt |
| Customs tariff number | 8419 8998 |
| Dimensions approx. incl. carton | 650 (W) X 570 (D) X 800 (H) mm |
| Net weight | 35 Kg |
| Gross weight carton | 45 Kg |